Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
Published on: February 7, 2014
Importance of Considering Temporal Variations in Pulse Wave Velocity for Accurate Blood Pressure Prediction
Aditya Satishkumar Bantwal1, Amit Kumar Bhayadia1, Hui Meng2
1Department of Mechanical and Aerospace Engineering, University at Buffalo, Buffalo, NY, 14260, USA.
Purpose:
Continuous, cuffless blood pressure (BP) monitoring devices based on measuring pulse wave velocity (PWV) or pulse transit time (PTT) are emerging but are often plagued by large prediction errors. A key issue is that these techniques typically rely on a single PWV value, assuming a linear response and small arterial wall deformations. However, arterial response to BP is inherently nonlinear, with PWV varying over time [PWV(t)] by up to 50% during a cardiac cycle. This study evaluates the impact of assuming a single PWV on BP prediction accuracy.
Method:
Using a Fluid-structure Interaction (FSI) testbed, we simulate the radial and common carotid arteries with the Holzapfel-Gasser-Ogden (HGO) constitutive model to capture nonlinear arterial behavior under a pulsatile physiological blood flow. Pressure data from FSI simulation are used as the ground truth, while inner area A(t) and two PWV values, at diastole and systole, serve as inputs to BP prediction models. Two models are tested: one using a single PWV value, emulating existing PWV-based BP prediction methods; another using the two PWV values to account for PWV(t).
Results:
The single-PWV BP model produced prediction errors of 17.44 mmHg and 6.57 mmHg for the radial and carotid arteries, respectively. The model incorporating two PWV values reduced these errors by 90.6% and 96.8%, respectively.
Conclusion:
Relying on a single PWV in BP prediction models can lead to significant errors. To improve BP accuracy, future efforts should focus on incorporating PWV(t), or at least both diastolic and systolic PWV values, into these models.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Pulse
The pulse serves as a clinical...
Special considerations while measuring pulse
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: