Spectral performance for iodine quantification of a dual-source, dual-kV photon counting detector CT

Tim Winfree1, Kevin Treb1, Cynthia McCollough1

  • 1Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.

Medical Physics
|February 10, 2025
PubMed
Summary

Dual-source photon counting detector CT (DS-PCD-CT) offers the lowest noise for iodine quantification, outperforming single-source PCD-CT and dual-source energy-integrating detector CT. This advancement improves material concentration accuracy in multi-energy CT imaging.

Related Concept Videos

Imaging Studies II: Positron Emission Tomography and Scintigraphy01:25

Imaging Studies II: Positron Emission Tomography and Scintigraphy

Positron Emission Tomography (PET) is a medical imaging technique that provides crucial insights into the body's physiological functions at a molecular level. It is an indispensable resource for diagnosing, staging, and monitoring various illnesses, notably cancer, neurological disorders, and cardiovascular conditions.
Fundamental Principles of PET
68
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation01:26

Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation

Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy  (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
183
UV–Vis Spectrometers01:14

UV–Vis Spectrometers

The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
1.3K
Atomic Absorption Spectroscopy: Instrumentation01:22

Atomic Absorption Spectroscopy: Instrumentation

An atomic absorption spectrophotometer (AAS) comprises several components: a radiation source, an atomizer, a monochromator, and a detector. The radiation source can be a hollow-cathode lamp (HCL) or an electrodeless-discharge lamp (EDL), both of which provide a narrow emission line of the required wavelength. However, some instruments use continuum sources and high-resolution monochromators to achieve a narrow range of radiation.
The atomizer used in AAS can be either a flame atomizer or an...
547