Related Experiment Video
Updated: May 28, 2025

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Patient-ventilator synchrony under non-invasive ventilation is improved by an automated real time waveform analysis
Yann Renaud1,2, Jocelyne Auroi3,4, Davy Cabrio5,3
1Adult Intensive Care Unit, Lausanne University Hospital, Lausanne, Switzerland. yann.renaud@chuv.ch.
Background:
Because of inherent leaks, obtaining good patient-ventilator synchrony during non-invasive ventilation (NIV) is challenging. The IntelliSync + ® software (Hamilton medical, Bonaduz, CH), that can be used together with the NIV mode, performs real-time automated analysis of airway pressure- and flow-time curves to detect the transition between inspiration and expiration. It then controls the ventilator inspiratory and expiratory valves to improve patient-ventilator synchrony. The main goal of this NIV bench study was to evaluate the impact of IntelliSync + ® on synchrony in the presence of leaks of 9 and 20 L/min in the tested ventilator circuit (no face mask used), with normal, obstructive and restrictive respiratory mechanics and two levels of NIV pressure support (PS 8 and 14 cmH2O). For this, the time needed to trigger the ventilator (Td) and the difference between the end of the simulated breath and the termination of pressurization (Tiex) were measured. The number of classical asynchronies and the ventilator pressurization capacity were also assessed.
Results:
Compared to NIV delivered with the classical NIV mode (compensating leaks and limiting inspiratory time to 2 s), activating IntelliSync + ® improved Tiex and, to a lesser extent, Td in clinically relevant setups. IntelliSync + ® also showed a trend towards reducing classical asynchronies, particularly directly after leak flow increase. The impact of the system was most significant with high PS levels and pathological respiratory mechanics. Especially, in the obstructive model, in the presence of large leak (20 L/min) and PS 14 cmH2O, Tiex decreased from 0.61 [0.56-0.64] to 0.16 [0.07-0.18] s and Td from 0.07 [0.06-0.08] to 0.06 [0.06-0.08] s. In less challenging situations, IntelliSync + ® was less beneficial. Overall, ventilator pressurization was improved when IntelliSync + ® was activated.
Conclusions:
In this NIV bench model, IntelliSync + ®, used in addition to NIV-PS, improved both expiratory and inspiratory synchrony. It was particularly efficient in the presence of obstructive and restrictive respiratory mechanics and high-pressure support levels. These pre-clinical results tend to support the ability of IntelliSync + ® to improve patient-ventilator synchrony in the presence of leaks and provide pre-clinical data supporting a clinical evaluation of the automated algorithm during NIV.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...

