Potential of photon-counting detector CT technology for contrast medium reduction in portal venous phase
Daniel Popp1,2, Martin Siedlecki1, Lena Friedrich1
1Department of Diagnostic and Interventional Radiology, University Hospital Augsburg, Faculty of Medicine, University of Augsburg, Augsburg, Germany.
Objectives:
To compare image quality and iodine attenuation intra-individually in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium (CM) volume.
Materials And Methods:
A prospectively acquired patient cohort between 04/2021 and 11/2023 was retrospectively screened if patients had the following combination of portal venous phase thoracoabdominal CT scans: (a) PCD-CT with 120 mL CM volume (PCD-CT120 mL), (b) PCD-CT with 100 mL CM volume (PCD-CT100 mL), and (c) prior energy-integrating detector CT (EID-CT) with 120 mL CM volume. On PCD-CT, virtual monoenergetic image (VMI) reconstructions at 70 keV were applied for both groups as well as additional VMI at 60 keV for PCD‑CT100 mL. Quantitative analyses including signal-to-noise (SNR) and contrast-to-noise ratios (CNR) and qualitative analyses were performed using a mixed linear effects model.
Results:
The final study cohort comprised 49 patients (mean age 67 [31-86] years, 12 female). Comparison to EID-CT was available in 33 patients. In standard 70 keV VMI reconstructions, PCD-CT100 mL was non-inferior to PCD-CT120 mL as well as to EID-CT120 mL for CNR in abdominal organs (all p > 0.050). The mixed linear effects model revealed significant differences between contrast volume groups for both contrast enhancement and image quality ratings. PCD-CT100 mL/70 keV demonstrated the smallest deviation from optimal contrast enhancement (-0.306, p < 0.001).
Conclusion:
In portal venous phase thoracoabdominal PCD-CT, a nearly 17% reduction in CM was achievable while maintaining subjective and objective image quality compared to prior higher CM volume PCD-CT scans within the same patients and still surpassing image quality of previous exams on an EID-CT system.
Key Points:
Question How do image quality and iodine attenuation intra-individually compare in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium volume. Findings PCD-CT scans exhibit superior quantitative and qualitative image quality compared to energy-integrating detector-CT acquisitions and are not negatively affected by contrast volume reductions up to 17%. Clinical relevance This study provides further evidence that PCD-CT enables a considerable reduction in iodine dose for portal venous phase acquisition, benefiting both patients and healthcare system costs.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...