Evaluating Carotid Plaque Stiffness with Ultrasound 2D Shear-Wave Elastography in Patients Undergoing Coronary Artery

Adel Alzahrani1, Amjad Ali Alharbi1, Amjad Khalid Alharbi1

  • 1Department of Diagnostic Radiology and Imaging, King Abdullah Medical City, Makkah 24246, Saudi Arabia.

PubMed

Insights

Ultrasound 2D shear-wave elastography (2D-SWE) reliably measures carotid plaque stiffness. Patients undergoing coronary artery bypass grafting (CABG) showed less stiff plaques, indicating potential for risk assessment.

Area of Science:

  • Cardiovascular imaging
  • Noninvasive atherosclerosis assessment
  • Biomechanical properties of arterial plaques

Background:

  • Coronary and carotid artery diseases are linked systemic atherosclerotic conditions.
  • Evaluating both coronary and carotid atherosclerosis is crucial in high-risk individuals.
  • Ultrasound 2D shear-wave elastography (2D-SWE) shows potential for noninvasive carotid plaque stiffness assessment.

Purpose of the Study:

  • To assess carotid plaque stiffness using 2D-SWE in patients undergoing coronary artery bypass grafting (CABG).
  • To compare carotid plaque stiffness between CABG patients and a control group.
  • To evaluate the reproducibility and correlation of 2D-SWE measurements with plaque characteristics.

Main Methods:

  • Prospective pilot study involving 32 patients (17 CABG, 15 controls) with 43 carotid plaques.
  • Bilateral carotid ultrasound with high-resolution linear transducer.
  • Carotid plaque stiffness quantified by 2D-SWE (SWV, YM); plaque characteristics including GSM quantified; intra-observer reproducibility assessed via ICCs.

Main Results:

  • Carotid plaques in CABG patients were significantly less stiff than controls (SWV: 3.64 m/s vs. 4.91 m/s; YM: 20.96 kPa vs. 72.54 kPa).
  • 2D-SWE measurements demonstrated excellent intra-observer reproducibility (SWV ICC=0.992, YM ICC=0.992).
  • A positive correlation was found between 2D-SWE (SWV, YM) and Gray Scale Median (GSM) values.

Conclusions:

  • Ultrasound 2D-SWE is a reliable tool for quantifying carotid plaque stiffness with high reproducibility.
  • Reduced carotid plaque stiffness in CABG patients suggests potential as a marker for high-risk plaques.
  • 2D-SWE may aid in assessing cerebrovascular risk in patients undergoing CABG.