Ultra-Thin Plastic Scintillator-Based Proton Detector for Timing Applications.
Mauricio Rodríguez Ramos1, Javier García López1,2, Michael Seimetz3
1Centro Nacional de Aceleradores (U. Sevilla, CSIC, J. de Andalucia), 41092 Seville, Spain.
Sensors (Basel, Switzerland)
|February 13, 2025
Summary
A new detector system using ultra-thin plastic scintillators was developed for precise charged particle timing. This system successfully achieved single ion detection, proving its utility as a diagnostic tool in accelerator applications.
Area of Science:
- Nuclear Instrumentation and Detectors
- Accelerator Physics and Technology
- Materials Science for Particle Detection
Background:
- Advanced detection systems are crucial for charged particle identification in laser-based accelerators.
- Precise time-of-flight (TOF) measurements are essential for analyzing particle trajectories and energies.
- Ultra-thin plastic scintillators offer low energy loss and minimal dispersion for MeV protons.
Purpose of the Study:
- To introduce and characterize a novel detection system for TOF and timing applications.
- To evaluate the performance of an ultra-thin EJ-214 plastic scintillator-based detector.
- To assess the detector's suitability as an external trigger and diagnostic tool at an ion beam nuclear microprobe.
Main Methods:
- Development of a detector system comprising an ultra-thin EJ-214 plastic scintillator, photomultiplier tube, and shielding.
- Characterization of temporal performance and ion transmission of the prototype detector.
- Analysis of scintillator thickness and uniformity using Rutherford backscattering spectrometry (RBS).
Main Results:
- Experimental scintillator thickness was found to deviate significantly (46%) from supplier specifications.
- Detector response to MeV protons showed position dependence but linearity with applied bias.
- Successful single ion detection was achieved, validating the system's diagnostic capabilities.
Conclusions:
- The developed detector system demonstrates effective single ion detection for timing applications.
- The findings highlight the importance of verifying material specifications in detector development.
- This system is a promising diagnostic tool for charged particle analysis in accelerator environments.