A Rare Case of Thrombus Aspiration Device Stuck in the Stent
Himanshu Gupta1, Palanivel Rajan1, Harkant Singh2
1Department of Cardiology, Post Graduate Institute of Medical Education and Research, Chandigarh, India.
Abstract:
Hardware loss and entrapment during percutaneous coronary intervention is one of the important complications, which an interventionist should anticipate in his practice. Basic knowledge about various extraction methods is essential to bail out in such situations. We describe a 74-year-old male, in whom a used thrombuster device got stuck in struts of an already deployed stent in the right coronary artery. The device could not be retrieved via various percutaneous maneuvers, necessitating an emergency surgical device retrieval and concomitant coronary bypass.