Performance of enzymatic creatinine methods in the pediatric concentration range
Hilda Hallander1, Magnus Lindén2,3, Catarina Lindqvist1
1Department of Clinical Chemistry, Halland Hospital, Varberg and Halmstad, Sweden.
This study evaluated creatinine measurement methods for children, finding varying accuracy in low ranges. Method bias correction needs careful attention for pediatric creatinine levels to ensure reliable kidney function estimates.
Area of Science:
- Clinical Chemistry
- Biomarker Analysis
- Pediatric Medicine
Background:
- Creatinine is a key biomarker for assessing kidney function and glomerular filtration rate in both adults and children.
- Recent studies evaluating creatinine measurement performance specifically within the pediatric concentration range are limited.
- Accurate creatinine measurement is crucial for timely diagnosis and management of kidney diseases in pediatric populations.
Purpose of the Study:
- To assess the measurement performance of commonly used enzymatic creatinine assays on four different platforms (Abbott Alinity, Radiometer ABL800, Roche Cobas, Siemens Atellica) within the pediatric concentration range.
- To compare these platforms against the gold standard isotope dilution mass spectrometry (IDMS) reference method.
- To evaluate the performance of Roche enzymatic methods using control sera from the Nordic Association of Clinical Chemistry.
Main Methods:
- Comparative analytical study involving four automated immunoassay platforms.
- Measurement of creatinine concentrations in pediatric-relevant samples.
- Comparison of results against the isotope dilution mass spectrometry (IDMS) reference method.
- Utilized Nordic Association of Clinical Chemistry control sera for secondary method comparison.
Main Results:
- Significant variability in accuracy was observed among the evaluated creatinine methods at low concentration ranges relevant to pediatrics.
- The relative differences between measurement platforms decreased as creatinine concentration increased, particularly below 75 µmol/L.
- Applying a single absolute factor for method bias correction, as sometimes recommended, may compromise measurement accuracy in the pediatric creatinine range.
Conclusions:
- Current enzymatic creatinine assays exhibit variable performance in the pediatric concentration range.
- Standardization and improved agreement between creatinine measurement methods are necessary for reliable pediatric kidney function assessment.
- Manufacturers and stakeholders must consider the unique challenges of low-concentration creatinine measurements in pediatric patients when developing bias correction strategies.
Related Concept Videos
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...
Determination of Renal Drug Clearance: Graphical and Midpoint Methods
The graphical method involves plotting the rate of drug excretion in urine against the plasma drug concentration. By analyzing the graph, the clearance can be calculated and obtained. Drugs rapidly excreted by the kidneys exhibit a...
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...