3D-printed template design to improve 125I seed plaque assembly accuracy for uveal brachytherapy
Related Experiment Video
Updated: May 27, 2025

Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
Kaitlyn Calabresi1, Jacqueline Emrich2, Sara Belko3
1Department of Radiation Oncology, Thomas Jefferson University Hospital, Philadelphia, PA 19107.
Purpose:
To demonstrate the utility of a QA program for seed localization, and to design a 3D-printed template to improve the accuracy of seed placement on custom-built 125I eye plaques for uveal brachytherapy.
Methods And Materials:
A seed localization analysis tool (SLAT) was developed in MATLAB to detect variations in seed placement relative to a treatment plan. A flexible seed placement template (3D-FSPT) was designed in CAD and printed using a Formlabs Form-3 3D printer. The 3D-FSPT and SLAT were tested using 3D-printed model eye plaques with nonradioactive seeds arranged following clinically-relevant treatment plans. Five clinical plaques were also evaluated.
Results:
SLAT detected submillimeter scale variations in seed position with 2.3% error relative to the plan's seed coordinates, and with an uncertainty of ± 0.01 mm. The average seed displacement on the model plaques with free-handed seed placement was 1.31 mm (SD = 0.61), and the average seed orientation difference was 5.27 degrees (SD = 4.77). The average seed displacement on the clinical plaques was 0.77 mm (SD = 0.42), and the average seed orientation difference was 4.41 degrees (SD = 3.00). For the clinical plaques, changes in dosimetry to the tumor apex and critical eye structures were within acceptable tolerances. Seed displacement (mm) (p < 0.001) and seed orientation differences (degrees) (p = 0.008) were significantly lower using the template to guide seed placement on the model plaques compared to free-handed seed placement.
Conclusions:
The feasibility of a 3D-FSPT and SLAT is demonstrated for improving seed placement accuracy relative to a treatment plan.