A protocol for the assessment of uroflowmeters
Alexandra Bacon1, Andrew Gammie1
1Bristol Urological Institute, Southmead Hospital, Bristol, UK.
Abstract:
Uroflowmetry plays an important role in the investigation of patients with lower urinary tract symptoms. We were required to assess a newly developed uroflowmeter. We thus aimed to produce a standardised protocol to test the accuracy and filtering of any new uroflowmeter. The accuracy of a newly developed uroflowmeter (Minze Uroflow®) was validated using a constant flow bottle and a cylindrical flow column. Two other machines were also tested alongside. We also assessed filtering by reproducing common artefacts in the laboratory. Finally, a questionnaire was constructed to assess the usability of the uroflowmeter by clinicians during a normal hospital flow clinic. A protocol to test new uroflowmeters was written and assessed. The protocol showed the following results for the tested uroflowmeters: a simple bench test using a constant flow bottle and cylindric column showed that the uroflow parameters (Qmax and Vvoid) were within the claimed accuracy range and ICS recommendations. The processing of the flow data by the systems effectively filtered noise, and the flow rate decline over the whole measurement range, as produced by the cylindrical flow column, was smooth and linear. Usability was assessed by clinicians in their routine clinical practice. The proposed tests meet the requirements of the ICS guidelines. We have designed a protocol which can be used by clinicians and researchers to validate the accuracy of their uroflowmeters, evaluate new models and ensure clinical usefulness.
Related Concept Videos
Pipe Flowrate Measurement
The orifice meter is a simple,...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Flow Table Test
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: