A novel automated IHC staining system for quality control application in ALK immunohistochemistry testing
Chunxiao Hou1, Xueru Song2, Hongwei Chen2
1Department of Pathology, Haining People's Hospital, Haining, Zhejiang, China.
A new automated quality control method for anaplastic lymphoma kinase (ALK) immunohistochemistry (IHC) in lung adenocarcinoma uses liquid controls. This method offers high accuracy and avoids using scarce patient tissue for reliable ALK testing.
Area of Science:
- Oncology
- Pathology
- Biotechnology
Background:
- Accurate anaplastic lymphoma kinase (ALK) testing is crucial for lung adenocarcinoma treatment.
- Established immunohistochemistry (IHC) methods require reliable positive and negative controls.
- Patient tissue controls for ALK IHC present ethical challenges and scarcity issues.
Purpose of the Study:
- To evaluate a novel automated solution for ALK IHC quality control management.
- To assess the performance of a new concentrated antibody (BP6165) on the LYNX480 PLUS platform.
- To compare automated liquid controls with traditional tissue controls.
Main Methods:
- Investigated a novel automated ALK IHC quality control system using BP6165 concentrated antibody on the LYNX480 PLUS platform.
- Compared the automated system with the established D5F3 antibody on the VENTANA system.
- Analyzed 87 lung adenocarcinoma specimens with known ALK status, re-verified by fluorescence in situ hybridization.
Main Results:
- The BP6165 antibody demonstrated high sensitivity (98.30%) and specificity (100%) in 87 biopsy specimens.
- Automated liquid controls (CLFs) showed improved morphology and cell distribution compared to manual application.
- Novel controls mimicked tissue control responses to varying antibody concentrations and antigen retrieval conditions.
Conclusions:
- The automated ALK IHC quality control solution provides a convenient and ethical alternative to scarce patient tissue.
- Standardized protocols with BP6165 antibody on the LYNX480 PLUS platform enhance reliable ALK testing for lung adenocarcinoma patients.
- This advancement facilitates wider laboratory adoption for determining eligibility for ALK tyrosine kinase inhibitor therapy.