Monte Carlo calculated beam quality correction factors for high energy photon and electron fields
Damian Czarnecki1, Klemens Zink2
1Institute of Medical Physics and Radiation Protection, University of Applied Sciences (THM), Wiesenstraße 14, Giessen, 35390, Germany.
Purpose:
Although several studies provide data on reference dosimetry, ionization chambers are in clinical use worldwide, for which no or only a few correction factors for beam quality are available. The aim of this study was to calculate beam quality correction factors in electron and photon fields for four cylindrical ionization chamber, i.e. CC08, CC04, Razor™ Chamber and Razor™ Nano Chamber (all IBA, Dosimetry, Schwarzenbruck, Germany).
Method:
Monte Carlo simulations were performed with the EGSnrc code system to calculate the beam quality correction factor kQ. Therefore, a detailed Monte Carlo based model of the four ionization chambers were created to calculate the dose in the sensitive volume. Electron and photon spectra of clinical linear accelerators within the entire energy range of therapeutic linear accelerators were used as radiation source.
Results:
kQ values as a function of the beam quality specifiers TPR20,10 and R50 were calculated for high energy photon and electron fields, respectively. The data was fitted against recommended equations for photon and electron dosimetry with a χ2 between 6.4e-7 and 4.2e-6.
Conclusion:
Electron and photon beam quality correction factors kQ were calculated by Monte Carlo simulations for the cylindrical ionization chambers CC08, CC04, Razor™ Chamber and Razor™ Nano Chamber to provide kQ values for absolute dosimetry according to the TRS 398 dosimetry protocol.
More Related Videos
08:31Sample Preparation and Experimental Design for In Situ Multi-Beam Transmission Electron Microscopy Irradiation Experiments
Published on: June 27, 2022
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
Related Concept Videos
Transmission Electron Microscopy
Thomson's e/m Experiment
A particle with charge q, speed v, and mass m enters an area from the top, where the magnetic and electric fields are perpendicular both to the particle's motion and to one another. The...
Overview of Electron Microscopy
Biological Effects of Radiation
NMR Spectrometers: Resolution and Error Correction