Intraocular Pressure Measurements With the Tono-Vera: A New Handheld Rebound Tonometer
Luiz A F Beniz1,2, Alessandro A Jammal1, Davina A Malek1
1Bascom Palmer Eye Institute, University of Miami, Miami, FL.
PréCis:
In a cross-sectional study comparing the novel Tono-Vera handheld rebound tonometer with the Ocular Response Analyzer, a strong agreement was found between the intraocular pressure measurements of the 2 devices.
Purpose:
To compare intraocular pressure (IOP) measurements obtained by the new Tono-Vera (Reichert Technologies), a novel handheld rebound tonometer with a proprietary camera-based positioning system, with those from the Ocular Response Analyzer (ORA, Reichert Technologies) non-contact tonometer.
Patients:
Fifty-four eyes of normal individuals and open angle glaucoma subjects.
Methods:
All subjects had at least 6 reliable measurements with the Tono-Vera and 3 with the ORA, acquired by the same technician in the same visit. The sample was randomly split into calibration and validation samples. An ordinary least squares linear regression model generated a calibration formula for Tono-Vera readings using ORA's Goldmann-correlated IOP (IOPg) as a target. Bland-Altman plots evaluated the agreement between Tono-Vera calibrated IOP (IOPtv) and ORA IOPg in the validation sample.
Main Outcome Measures:
Agreement between IOPtv and IOPg measurements.
Results:
Fifty glaucoma and 4 normal eyes were included, mean age 68.3 ± 9.4 years, mean central corneal thickness 535.0 ± 39.6 μm. In the validation sample, mean IOPtv was 14.6 ± 5.2 mmHg (range, 7.4-28.5 mmHg) versus 15.0 ± 5.3 mmHg (range, 6.6-28.6 mmHg) for ORA IOPg ( R ² = 0.782). Bland-Altman plots showed a mean difference of -0.4 ± 2.5 mmHg (95% limits of agreement: -5.4 to 4.5 mmHg). 58.6%, 79.3%, and 96.5% of IOPtv measurements fell within ± 2, ± 3, and ± 5 mmHg of ORA measurements. No fixed or proportional biases were observed ( P = 0.620).
Conclusion:
There was a strong correlation and agreement in IOP measurements between calibrated Tono-Vera IOP and ORA IOPg.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Glaucoma: Overview
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...