Related Experiment Video
Updated: May 24, 2025

A Novel Approach for Documenting Phosphenes Induced by Transcranial Magnetic Stimulation
Published on: April 1, 2010
A dedicated RF coil design for phosphorus MRS in the occipital lobe
Abstract:
The dedicated RF coil for phosphorus magnetic resonance spectrum (31P MRS) was designed as a double-channel transceiver integrated surface coil structure. It features a semi-curved "pillow" shape, specifically tailored to capture phosphorus magnetic resonance spectroscopy signals in the occipital lobe of the human brain at high magnetic field strength (3T). Through RF magnetic simulation and bench tests, it has been demonstrated that the two channels of the coil exhibit excellent resonance and matching properties at the 3T 31P NMR frequency. Additionally, the mutual perpendicular RF magnetic field configuration ensures that the coupling between the channels is negligible. When conducting 31P MRS scans with a phosphorus phantom, it was observed that the spectral lines displayed a smooth baseline, prominent peaks, and an exceptional signal-to-noise ratio (SNR) after 2-3 calibrations. These results indicate the efficacy of the coil in capturing high- quality 31P MRS signals, offering valuable insights into the metabolic activity within the occipital lobe.
Related Concept Videos
Atomic Nuclei: Magnetic Resonance
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
NMR Spectrometers: Resolution and Error Correction
Magnetic Resonance Imaging
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...