Can Bioimpedance Analysis Be Used to Estimate Absolute Blood Volume in Hemodialysis Patients?
Joachim Kron1, John Volkenandt2, Stefanie Broszeit2
1KfH Kidney Center Berlin-Köpenick, Berlin, Germany.
Introduction:
Under physiological conditions, blood volume and extracellular volume are in a ratio of 1-3, even in hemodialysis patients. The question therefore arises: can blood volume be inferred from the bioimpedance analysis data? The aim of the study was to compare the blood volumes calculated from extracellular volume determined by bioimpedance analysis data to the actually measured blood volumes.
Methods:
Immediately before treatment, extracellular volume and volume overload were evaluated by bioimpedance spectroscopy. The actual blood volume was determined by indicator dilution, using an on-line infusate bolus and subsequent calculation with the data from the relative blood volume monitor. Alternatively, blood volume was calculated from extracellular volume divided by 3 and compared to the measured blood volume.
Findings:
Overall, there were no significant differences between measured (5.56 ± 1.47 L) and calculated (5.79 ± 1.30 L) blood volumes. However, intra-individually, there were very large discrepancies with a range of -1.409 to 1.450 L. Median absolute deviation was 382 mL corresponding to 6.2 mL/kg. The differences between measured and calculated blood volumes correlated significantly (r = -0.98; p < 0.001) with the blood to extracellular volume ratio.
Discussion:
In almost half of patients, blood volume can be inferred from bioimpedance data with sufficient certainty. But the greater the deviation from the physiological blood to extracellular volume ratio of 1-3, the more the calculated blood volumes differ from the measured values. For this reason, bioimpedance data should not be used uncritically to set the ultrafiltration.
More Related Videos
10:21Evaluation of Hydration Status by Bioelectrical Impedance Vector Analysis in Patients with Ischemic Heart Disease Undergoing Exercise Stress Test
Published on: September 22, 2023
04:05Author Spotlight: Unveiling Prognostic Indicators in Heart Failure - The Role of Phase Angle and Bioelectrical Impedance Analysis
Published on: June 30, 2023
Related Concept Videos
Dialysis
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...