Clinical photon-counting CT increases CT number precision and reduces patient size dependence compared to single- and
Jessica D Flores1,2, Gavin Poludniowski2,3, Adrian Szum1,4
1Department of Clinical Neuroscience, Karolinska Institutet, Stockholm, 141 86, Sweden.
Objectives:
To study whether photon-counting computed tomography (PCCT) can improve CT number accuracy and precision and reduce patient size dependence compared to dual-energy CT (DECT) virtual monoenergetic imaging (VMI) and single-energy CT (SECT).
Methods:
Clinical PCCT, DECT, and SECT scanners were used to image a multi-energy quality assurance phantom and tissue-equivalent inserts with/without an outer nested annulus, representing 2 object sizes (18 and 33 cm). CT numbers were converted to linear attenuation coefficients (LAC) and regions of interest applied. Theoretical monoenergetic LAC were calculated from known elemental compositions as a ground truth. Percent differences in mean LAC between phantom sizes, between mean and theoretical LAC, and its coefficient of variation (COV) were calculated.
Results:
Mean LAC percent differences between small and larger phantoms were highest in DECT (within -3% to 9%) and SECT (within 1%-5%), particularly at higher calcium and iodine concentrations, while being relatively constant in PCCT over material concentrations and VMI energies (within ±2%). The COV in mean LAC was consistently lower (about 2-5 times) in PCCT relative to DECT and SECT for calcium in the large phantom. With consideration of the theoretical uncertainties of 2%, both PCCT and DECT showed comparable agreement to theoretical LAC.
Conclusions:
PCCT VMI produces CT numbers with less dependence on patient size and increased precision in large object sizes than DECT VMI and SECT.
Advances In Knowledge:
Clinical PCCT provides less variable CT numbers than DECT and SECT with less sensitivity to the imaged object size.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...