Related Experiment Videos
Catheters for arterial pressure monitoring in pediatrics
Critical Care Medicine
|July 1, 1985
Abstract:
To evaluate the accuracy of an underdamped catheter transducer recorder used to measure arterial blood pressure, natural frequency and damping ratio were determined by transient testing with two types of 22-ga and 24-ga catheters. Damping coefficients were considerably higher and resonant frequencies lower with the 24-ga catheters. The difference between waveforms of known pressure and HR and the output waveforms produced by the catheter transducer systems was determined by using a BP systems calibrator. Catheter size did not affect the difference in recorded pressures.