Related Experiment Video
Updated: May 22, 2025

A Microfluidic-based Electrochemical Biochip for Label-free DNA Hybridization Analysis
Published on: September 10, 2014
Distance-Readout Paper-Based Microfluidic Chip with a DNA Hydrogel Valve for AFB1 Detection
Peng Chen1, Wenhao Han2, Yansheng Li2,3
1Beijing Key Laboratory of Traditional Chinese Medicine Basic Research on Prevention and Treatment for Major Diseases, Robot Intelligent Laboratory of Traditional Chinese Medicine, Experimental Research Center, China Academy of Chinese Medical Sciences, Beijing 100700, P. R. China.
Abstract:
Accurate and rapid aflatoxin B1 (AFB1) detection is essential for ensuring the safety of food supplies. In this paper, we introduce a distance-readout paper-based microfluidic chip (DPMC) that offers a sensitive and reliable method for the detection of AFB1. The DPMC comprises a DNA hydrogel sensitive valve and a paper-based capillary channel. Upon exposure to AFB1, the hydrogel valve regulates the flow speed of the tested liquid into the capillary channel. Quantitative detection of AFB1 can be achieved without the need for complex instrumentation for external equipment by visually observing the distance traveled by the tested liquid through the capillary channel over a specified period. Under optimal conditions, the DPMC allows for quantitative detection of AFB1 solution concentrations ranging from 100 to 1000 pM by the naked eye, with a detection limit of 17.64 pM. This method has been successfully employed for quantitative detection of AFB1 in lotus seed samples, yielding a recovery rate between 85.2% and 118.4%. This approach provides a rapid, portable, sensitive, and highly selective visualization platform for on-site AFB1 detection.