Facet joint loading after 1-, 2- and 3-level cervical disc arthroplasty: a comparison of spiked versus keeled
Jason M Cuellar1, Todd Lanman1, Nicole Mottole2
1Cedars-Sinai Spine Center, Los Angeles, CA, United States.
Background:
The purpose of this study was to examine facet contact forces above, below, and at surgical index levels induced by artificial disc implantation and compare the results from spiked versus keeled baseplates comprising the arthroplasty device.
Methods:
Human specimens from C2 to C7 were subjected to flexion, extension, and lateral bending prior to, and following random allocation to spiked or keeled cervical arthroplasty at the index (C5-C6), inferior (C6-C7), and superior (C4-C5) levels. Thin film force sensors were inserted unilaterally into the corresponding facets prior to intact testing. Force data was normalized to the minimum forces recorded during each loading mode under each condition, reported as (Max/Min) force ratio and subjected to a 1-way ANOVA with Dunnett's post-hoc tests for comparison to intact specimens.
Results:
Under flexion, compared to intact, all levels displayed a significant reduction in force ratio following a 1- and 3-level implantation for the spiked baseplate device. An increase in force ratio was observed at the index level for a 2-level implantation but was mitigated with the completion of a superior device insertion. No statistical differences were noted for keeled devices. In extension, the spiked baseplate device reduced the force ratio for 1- and 2-level implantations. A 3-level insertion did not alter facet force ratios. For the keeled device, no statistical changes were noted. Lateral bending associated with spiked devices resulted in statistically reduced or nonsignificant changes in facet loading ratios. The keeled devices did not display significant changes to facet force ratios.
Conclusions:
Implantation of multilevel disc devices can reduce or sustain unaltered facet loading conditions. In general, 3-level arthroplasty statistically reduced or does not increase facet force ratios compared to intact values. The use of spiked versus keel device baseplates is a clinical selection involving anterior/posterior placement and endplate degeneration conditions.
More Related Videos
07:33In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
Structural Joints: Cartilaginous Joints
There are two types of cartilaginous joints:
Synchondrosis
A synchondrosis ("joined by cartilage") is a cartilaginous joint where bones are connected by hyaline cartilage. Synchondrosis may be temporary or...
Degenerative Disc Disease I: Introduction
Degenerative Disc Disease ll: Pathophysiology