Evaluation of In-Ear and Fingertip-Based Photoplethysmography Sensors for Measuring Cardiac Vagal Tone Relevant Heart
Ankit Parikh1,2, Gwyn Lewis3, Hamid GholamHosseini4
1School of Clinical Sciences, Auckland University of Technology, Auckland 0627, New Zealand.
Abstract:
This paper presents a study undertaken to evaluate the sensor systems that were shortlisted to be used in the development of a portable respiratory-gated transcutaneous auricular vagus nerve stimulation (taVNS) system. To date, all published studies assessing respiratory-gated taVNS have been performed in controlled laboratory environments. This limitation arises from the reliance on non-portable sensing equipment, which poses significant logistical challenges. Therefore, we recognised a need to develop a portable sensor system for future research, enabling participants to perform respiratory-gated stimulation conveniently from their homes. This study aimed to measure the accuracy of an in-ear and a fingertip-based photoplethysmography (PPG) sensor in measuring cardiac vagal tone relevant heart rate variability (HRV) parameters of root mean square of successive R-R interval differences (RMSSDs) and the high-frequency (HF) component of HRV. Thirty healthy participants wore the prototype sensor equipment and the gold standard electrocardiogram (ECG) equipment to record beat-to-beat intervals simultaneously during 10 min of normal breathing and 10 min of deep slow breathing (DSB). Additionally, a stretch sensor was evaluated to measure its accuracy in detecting exhalation when compared to the gold standard sensor. We used Bland-Altman analysis to establish the agreement between the prototypes and the ECG system. Intraclass correlation coefficients (ICCs) were calculated to establish consistency between the prototypes and the ECG system. For the stretch sensor, the true positive rate (TPR), false positive rate (FPR), and false negative rate (FNR) were calculated. Results indicate that while ICC values were generally good to excellent, only the fingertip-based sensor had an acceptable level of agreement in measuring RMSSDs during both breathing phases. Only the fingertip-based sensor had an acceptable level of agreement during normal breathing in measuring HF-HRV. The study highlights that a high correlation between sensors does not necessarily translate into a high level of agreement. In the case of the stretch sensor, it had an acceptable level of accuracy with a mean TPR of 85% during normal breathing and 95% during DSB. The results show that the fingertip-based sensor and the stretch sensor had acceptable levels of accuracy for use in the development of the respiratory-gated taVNS system.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...