Related Experiment Video
Updated: May 6, 2026

Imaging Membrane Potential with Two Types of Genetically Encoded Fluorescent Voltage Sensors
Published on: February 4, 2016
ElectroFluor Voltage-Sensitive Dyes: Comprehensive Analysis of Wavelength-Dependent Sensitivity and Cross-Channel
Srdjan D Antic1,2, Ping Yan3, Corey D Acker3
1Neuroscience, UConn Health, School of Medicine, Institute for Systems Genomics, Farmington, Connecticut, USA.
Abstract:
New voltage-sensitive ElectroFluor (EF) dyes that emit across the visible and near-infrared spectrum (e.g., 730 nm) were recently developed. We evaluated EF-530, EF-630, and EF-730p-dyes spectrally orthogonal to green fluorescent protein (GFP)-at excitation wavelengths outside the conventional 470 nm range used for GFP-based indicators. Although previously applied in cardiac voltage imaging, their performance in neuronal tissue remains untested. We performed side-by-side comparisons using population voltage imaging in mouse cerebral cortex slices at optimal excitation wavelengths (530, 630, and 730 nm) and assessed cross-channel signal bleed-through across four excitation wavelengths (475, 530, 630, and 730 nm). All dyes produced robust optical signals at their optimal wavelengths, though non-preferred channels exhibited bleed-through with distinct amplitudes, polarities, and photobleaching patterns. These results provide detailed quantifications of EF dye performance for neuronal population imaging.
Related Concept Videos
Voltmeter
An ideal voltmeter would have infinite resistance, so connecting it between two points in a circuit would not alter any of the currents. Real voltmeters always have...
Capillary Electrophoresis: Instrumentation