Anterior variable-angle locked plating of complex patella fractures - Does the type of polar screw matter?
Karl Stoffel1, Ivan Zderic2, Torsten Pastor3
1University Hospital Basel, Petersgraben 4, 4031 Basel, Switzerland.
Background:
Treatment of complex patella fractures represents a challenging clinical problem. Whereas controversy exists regarding the most appropriate fixation method, anterior plating is an emerging treatment option. A novel design of anterior variable-angle locking plates features radiating legs with integrated screw holes aiming at capturing fragments of the distal comminuted zone via either a locking or a cortical polar screw. The aim of this study was to compare the biomechanical performance of the novel anterior variable-angle locking plates used for fixation of complex patella fractures with either a locking or a cortical polar screw.
Methods:
Complex five-part AO/OTA 34-C3 patella fractures mimicking comminution around the distal pole were simulated in twelve human cadaveric knees. Specimens were randomized to two groups for treatment with an anterior variable-angle locking plate fixing the patella with either one locking or one cortical caudo-cranial bicortical polar screw. Biomechanical testing was performed over 5000 cycles by pulling on the quadriceps tendon, simulating active knee extension and passive flexion within the range from 90° flexion to full extension. Interfragmentary movements were captured by motion tracking.
Findings:
No significant differences between the two fixation techniques were detected in terms of longitudinal and shear articular displacements, and relative rotations around the mediolateral axis measured between the proximal and distal fragments at the central patella aspect between 1000 and 5000 cycles, P ≥ 0.116.
Interpretation:
From a biomechanical perspective, anterior locked plating of complex patella fractures provides comparable fixation stability when using either a locking or a cortical polar screw.
More Related Videos
05:34A Mini-Invasive Internal Fixation Technique for Studying Immobilization-Induced Knee Flexion Contracture in Rats
Published on: May 20, 2019
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Angle of Twist - Elastic Range
Angle of Twist: Problem Solving
Torsion of Noncircular Members
Unsymmetric Bending - Angle of Neutral Axis
When a bending moment is applied at an angle θ concerning the vertical axis of a symmetrical member, it can be resolved into components along the member's principal centroidal axes. The...
Stability of structures