Orthodontic appliances and their diagnostic impact to brain MRI
Lisa Latzko1, Anna Schmit2, Bernhard Glodny2
1Department of Dental and Oral Medicine and Cranio-Maxillofacial and Oral Surgery, Medical University of Innsbruck, Innsbruck, Austria.
Objective:
The aim of this study was to display and quantify signal loss artifacts in 1.5T and 3T brain MRI on a volunteer with different orthodontic appliances.
Materials And Methods:
In this experimental study, three different orthodontic appliances were examined on a 1.5T and a 3T MRI scanner in a healthy adult with normal dental occlusion: stainless-steel brackets paired with a nickel-titanium archwire; brackets, archwire, and stainless-steel molar bands; brackets, archwire, molar bands, and a stainless-steel trans-palatal archwire. Assessment of diverse anatomical structures, including different cerebral structures and blood vessels, was conducted using a six-point Likert scale.
Results:
Utilizing conventional stainless-steel brackets and a nickel-titanium archwire, with or without the inclusion of stainless-steel molar bands, all cerebral structures demonstrated satisfactory assessability with high diagnostic quality under both 1.5T and 3T MRI. For example, with an average rating of 85/85 for T2 and 77/85 for susceptibility-weighted imaging (SWI). Upon introduction of the stainless-steel trans-palatal archwire, additional artifacts were observed, predominantly manifesting in SWI (20/85), diffusion-weighted imaging (DWI) sequences (31/85), and phase contrast angiography (PCA) (17/20). Differences in artifact severity were mainly observed in the SWI and DWI sequences.
Conclusion:
Based on the findings of this study, it is not imperative to entirely remove orthodontic appliances to achieve sufficient diagnostic quality in brain MRI. In instances where SWI or DWI sequences are necessitated, the removal of solely the trans-palatal stainless-steel archwire should be contemplated, given its straightforward execution.
Clinical Relevance:
These results highlight the potential to reduce injury risk during orthodontic appliance removal, expedite imaging procedures, and consequently accelerate diagnostic processes, particularly crucial in emergencies.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...