1Department of Chemistry and Biochemistry, Brigham Young University, 84602, Provo, UT, USA.
This study presents a novel method for fabricating versatile paper-based analytical devices (μPADs) for rapid lactate quantification using 3D-printed electrodes and Prussian Blue. The developed system requires minimal sample volume and demonstrates applicability in real sweat analysis.
Area of Science:
- Electrochemistry
- Materials Science
- Analytical Chemistry
Background:
- Paper-based analytical devices (μPADs) offer low-cost, portable sensing solutions.
- Efficient fabrication of integrated electrode systems for μPADs remains a challenge.
- Lactate detection is crucial for various biomedical and sports monitoring applications.
Purpose of the Study:
- To develop a straightforward method for fabricating 3D-printed electrodes integrated into μPADs for lactate quantification.
- To optimize the μPAD system for rapid, sensitive, and low-volume sample analysis.
- To demonstrate the analytical applicability of the fabricated μPADs for lactate determination in real sweat samples.
Main Methods:
- Fabrication of poly(methyl methacrylate) (PMMA) molds using a 3D printing pen and carbon black filament.
- Modification of working electrodes with Prussian Blue (PB), lactate oxidase (LOx)/chitosan, and Nafion.
- Integration of a 3-electrode thermoplastic chip with μPADs, utilizing radial flow for sample analysis.
- Characterization of μPADs using food dye, ferricyanide, and optimization of paper substrate pore size and injection volume.
Main Results:
- The integrated μPAD system demonstrated rapid responses with sharp, intense transient signals.
- Optimized μPADs exhibited a linear range of 0.5 to 4 mmol L⁻¹ for lactate determination.
- The system successfully quantified lactate in real sweat samples using only 2 μL of sample.
Conclusions:
- A facile approach for fabricating integrated 3D-printed electrodes on μPADs was successfully developed.
- The developed μPADs offer a promising platform for rapid, low-volume lactate quantification.
- This technology opens avenues for new sensing applications requiring minimal sample volumes.