Increased VA-ECMO Pump Speed Reduces Left Atrial Pressure: Insights from a Novel Biventricular Heart Model
Anirudhan Kasavaraj1, Christian Said1,2, Laurence Antony Boss1
1Faculty of Medicine, UNSW Sydney, Kensington 2052, Australia.
Background And Aims:
The effect of veno-arterial extracorporeal membrane oxygenation (VA-ECMO) on left atrial pressure (LAP) in the presence of interventricular interaction and the Frank-Starling mechanism is unknown. We developed and validated a mock circulatory loop (MCL) incorporating a novel, 3D-printed biventricular heart model and Frank-Starling algorithm, and used this model to assess the determinants of LAP during VA-ECMO support.
Methods:
The MCL was designed to allow a separate ventricle or biventricular configuration, with or without an active Frank-Starling mechanism. The biventricular model with Frank-Starling mechanism was validated in terms of (1) the presence and degree of ventricular interactions; (2) its ability to simulate Frank-Starling physiology; and (3) its capacity to simulate normal and pathological cardiac states. In the separate ventricle and biventricular with Frank-Starling models, we assessed the effect on LAP of changes in mean aortic pressure (mAoP), ECMO pump speed, LV contractility and ECMO return flow direction.
Results:
In the biventricular configuration, clamping RA inflow decreased RAP, with a concurrent decrease in LAP, consistent with direct ventricular interaction. With a programmed Frank-Starling mechanism, decreasing RAP was associated with a significant reduction in both LV outflow and LV end-systolic pressure. In the biventricular model with a Frank-Starling algorithm, the MCL was able to reproduce pre-defined normal and pathological cardiac output, and arterial and ventricular pressures. Increasing aortic pressure caused a linear increase in LAP in the separate ventricle model, which was attenuated in the biventricular model with Frank-Starling mechanism. Increasing ECMO pump speed caused no change in LAP in the separate ventricle model (p = 0.75), but significantly decreased LAP in the biventricular model with Frank-Starling mechanism (p = 0.039), with stabilization of LAP at the highest pump speeds. Changing the direction of VA-ECMO return flow did not affect LAP in either the separate ventricle (p = 0.91) or biventricular model with Frank-Starling mechanism (p = 0.76).
Conclusions:
Interventricular interactions and the Frank-Starling mechanism can be simulated in a physical, biventricular MCL. In their presence, the effects of VA-ECMO on LAP are mitigated, with LAP reduction and stabilization at maximal VA-ECMO speeds.