Integration of Group A Streptococcus Rapid Tests with the Open Fluidic CandyCollect Device
J Carlos Sanchez1, Ingrid H Robertson1, Victoria A M Shinkawa1
1Department of Chemistry, University of Washington, Seattle, Washington 98195, United States.
Analytical Chemistry
|April 2, 2025
Summary
The CandyCollect lollipop device offers a comfortable oral fluid collection for rapid antigen detection tests (RADT) targeting Group A Streptococcus (GAS). This novel method successfully detected GAS in pediatric samples, showing promise for improved diagnostics.
Area of Science:
- Biotechnology and Medical Diagnostics
- Microbiology and Infectious Diseases
Background:
- Traditional pharyngeal swabs for Group A Streptococcus (GAS) detection can be uncomfortable for users, particularly children.
- Rapid antigen detection tests (RADTs) offer quick results but rely on effective sample collection methods.
Purpose of the Study:
- To evaluate the CandyCollect, a lollipop-inspired oral fluid device, as a user-friendly alternative for sample collection in GAS RADTs.
- To optimize a protocol for integrating the CandyCollect device with existing RADT kits for GAS detection.
Main Methods:
- Developed and optimized an elution protocol for oral fluid samples collected with the CandyCollect device for use with RADT kits.
- Conducted *in vitro* experiments using spiked saliva to determine optimal reagents and elution volumes for RADT readout.
- Validated the optimized protocol using clinical samples from pediatric patients diagnosed with GAS pharyngitis.
Main Results:
- The CandyCollect device coupled with RADT successfully detected *Streptococcus pyogenes* at concentrations as low as 1.5 × 105 CFU/mL.
- Image analysis of lateral flow assay (LFA) strips provided objective quantification of test results.
- Four out of six pediatric patients with confirmed GAS pharyngitis showed positive results using the CandyCollect-integrated RADT.
Conclusions:
- The CandyCollect device is a viable and comfortable alternative for oral fluid sample collection in GAS RADTs.
- The optimized protocol demonstrates the potential of CandyCollect for accurate and user-friendly detection of GAS infections in clinical settings.