Leaf Spray Mass Spectrometry: A Rapid Ambient Ionization Technique to Directly Assess Metabolites from Plant Tissues
Published on: June 21, 2018
Internal Standard Utilization Strategies for Quantitative Paper Spray Mass Spectrometry
Lucas R Abruzzi1,2, John-Clare Laxton1,2, Taelor M Zarkovic1,2
1Department of Chemistry, Vancouver Island University, Nanaimo, BC V9R 5S5, Canada.
Optimizing internal standards (ISTD) for paper spray mass spectrometry (PS-MS) is crucial. Predepositing ISTD, especially with automation, significantly improves precision and accuracy for quantitative analysis, meeting forensic science standards.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
- Forensic Science
Background:
- Premixing internal standards (ISTD) with samples for paper spray mass spectrometry (PS-MS) is inefficient and not universally applicable.
- Independent ISTD deposition on paper, while feasible, can compromise quantitative accuracy and precision.
Purpose of the Study:
- To evaluate different ISTD utilization strategies for PS-MS to identify sources of irreproducibility and bias.
- To compare the quantitative performance of premixed, predeposited, and postdeposited ISTD methods.
- To assess the impact of robotic versus manual pipetting on ISTD deposition accuracy.
Main Methods:
- Investigated ISTD deposition strategies (premixed, predeposited, postdeposited) using drugs of misuse as analytes.
- Assessed performance based on precision (CV) and accuracy (bias) with varying ISTD volumes.
- Compared robotic liquid handling systems with manual pipetting for ISTD deposition.
- Utilized strategy-matched calibrations to correct for systematic biases.
Main Results:
- Premixed ISTD yielded the best results (average CV = 1%, |bias| = 5%), significantly outperforming independently deposited ISTD (average CV = 18%, |bias| = 61%).
- Predeposited ISTD with a robotic handler showed optimal performance at lower volumes (CV = 8% for 2 μL).
- Manual pipetting of low-volume predeposited ISTD resulted in poor performance, while postdeposited ISTD was generally inferior.
Conclusions:
- Predeposition of ISTD, particularly using automated systems, enhances quantitative performance in PS-MS.
- Strategy-matched calibration effectively corrects biases across different ISTD utilization methods.
- Achieved American Academy of Forensic Sciences (AAFS) accuracy and precision requirements in most evaluated scenarios.
More Related Videos
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview