Construct stability in anterior open wedge osteotomy: Assessing 3D-printed polylactic acid wedges against cortical
Kai Hoffeld1, Jan P Hockmann1, Christopher Wahlers1
1University of Cologne, Faculty of Medicine and University Hospital of Cologne, Department of Orthopaedic, Trauma and Plastic Surgery, Germany.
Background:
This in-vitro biomechanical study aimed to assess whether using a wedge as a gap filler in anterior open wedge osteotomy of the proximal tibia reduces load on the screw-plate construct and to evaluate if a 3D-printed polylactic acid wedge offers similar biomechanical stability like a bone wedge.
Methods:
Twenty-four cadaveric proximal tibiae were randomly assigned to one of three groups: TomoFix plate alone as control group, TomoFix plate with a bone wedge, and TomoFix plate with a 3D-printed polylactic acid wedge. The specimens were subjected to axial loads of 400 N, 800 N, and 1200 N for 1000 cycles. Compression depth and axial stiffness were measured to assess stability.
Findings:
There were no significant differences in overall compression depth or axial stiffness between the groups (p > 0.05). At 1200 N, hinge fractures occurred in 6 of 8 specimens in the control group, compared to 0 of 8 in the wedge groups. Both bone and polylactic acid wedges improved construct stability, with the polylactic acid wedge performing comparably to the bone wedge.
Interpretation:
Gap fillers, whether bone or polylactic acid wedges, may contribute construct stability in anterior open wedge osteotomy of the proximal tibia. The 3D-printed polylactic acid wedge may be biomechanically comparable to the bone wedge, suggesting its potential as an alternative in clinical applications.