Metal-on-Metal Total Hip Arthroplasties: Why Do They Fail?
T Kyle Stoops1, Josué G Layuno-Matos2, Peter Simon2
1Indiana Joint Replacement Institute, Noblesville, IN, United States.
Metal sensitivity reactions are not the main cause of painful failed metal-on-metal hip replacements. Large fluid collections and high cobalt levels predict these reactions, but many painful cases have no clear cause.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Toxicology
Background:
- Metal-on-metal (MoM) total hip arthroplasty (THA) can fail due to metal sensitivity reactions (MSRs).
- However, MSRs may not be the primary reason for failure in all painful MoM THA cases.
- This study investigates the diverse failure modes of revised MoM THAs.
Purpose of the Study:
- To determine the failure etiologies of painful revised MoM THAs.
- To analyze operative indications and clinical outcomes following revision.
- To identify predictors of metal sensitivity reactions in MoM THA failure.
Main Methods:
- Review of 90 consecutive revised MoM THAs.
- Preoperative assessment including inflammatory markers, metal ion levels, imaging (MRI), and synovial fluid analysis.
- Evaluation of operative histopathology and clinical outcomes post-revision.
Main Results:
- Metal sensitivity reactions (MSRs) accounted for 36% of MoM failures; 64% were due to non-MSR etiologies like infection, loosening, and mechanical issues.
- Large periarticular fluid collections (OR 19.2) and elevated cobalt (Co) levels were highly predictive of MSR.
- 18% of painful MoM THAs had no identifiable cause, with 69% experiencing persistent pain after revision.
Conclusions:
- The majority of revised MoM THAs fail due to reasons other than MSR.
- Periarticular fluid collections and elevated Co levels are strong indicators of MSR.
- Unexplained painful MoM THA poses a significant challenge, often leading to persistent pain post-revision.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
09:08Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Related Concept Videos
Bonding in Metals
Fatigue
Imperfections in Crystal Structure: Non-Stoichiometric Defects