Real-World Performance of Personal Continuous Glucose Monitors During Hospitalization
Adrian G Dumitrascu1, Michelle Perry2, Ana-Maria Chindris2
1Division of Hospital Internal Medicine, Mayo Clinic Florida, Jacksonville, Florida, USA.
Diabetes Technology & Therapeutics
|April 14, 2025
Summary
Patients' personal continuous glucose monitors (CGMs) showed adequate hospital accuracy, except in hypoglycemia. Vital signs and lab values did not affect CGM performance, indicating reliable inpatient glucose monitoring.
Area of Science:
- Endocrinology
- Medical Devices
- Diabetes Management
Background:
- Continuous glucose monitors (CGMs) are increasingly used for diabetes management.
- Assessing CGM accuracy in inpatient settings is crucial for patient safety and effective glycemic control.
Purpose of the Study:
- To evaluate the accuracy of personal CGMs used by patients in a hospital setting.
- To determine the correlation between CGM accuracy and laboratory glucose values, as well as vital signs.
- To compare inpatient CGM performance metrics with published literature and clinical guidelines.
Main Methods:
- Adult patients with diabetes using personal CGMs upon admission were enrolled in a noncritical care setting.
- CGM readings were time-paired with point-of-care (POC) glucometers and laboratory (Lab) blood glucose measurements.
- Accuracy was assessed using mean absolute relative difference (MARD) and Clarke Error Grids; correlation analysis was performed for vital signs and lab values.
Main Results:
- CGM-POC pairs yielded a MARD of 13.7% (Dexcom: 14.4%, Libre: 11.8%), and CGM-Lab pairs showed a MARD of 13.6% (Dexcom: 12.7%, Libre: 15.4%).
- Over 98% of CGM-POC and CGM-Lab pairs fell within acceptable zones (A and B) on Clarke Error Grids.
- No significant correlation was found between CGM accuracy and mean arterial blood pressure, hemoglobin, glomerular filtration rate, or pulse oximetry.
Conclusions:
- Personal CGMs demonstrate adequate accuracy for inpatient glucose monitoring, with limitations noted in hypoglycemia ranges.
- Inpatient CGM performance is not significantly impacted by vital signs or laboratory results.
- Observed time-in-range (TIR) and glucose variability metrics align with inpatient guidelines and are comparable to hospital-inserted sensors.
Related Concept Videos
Pulse rhythm
721
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
721
Hypoglycemia and Glucagon
142
Without prolonged fasting, healthy individuals maintain blood glucose levels above 3.5 mM due to a well-adapted neuroendocrine counterregulatory system that effectively prevents acute hypoglycemia, a potentially life-threatening condition. The primary clinical scenarios for hypoglycemia encompass diabetes treatment, inappropriate production of endogenous insulin or insulin-like substances by tumors, and the use of glucose-lowering agents in non-diabetic individuals. Notably, hypoglycemia in the...
142
Errors occurring during blood pressure monitoring
568
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
568
SBAR II: Application of SBAR
4.2K
SBAR is an effective communication tool used by healthcare professionals to communicate patient information accurately. SBAR stands for Situation, Background, Assessment, and Recommendation. For a better understanding, an example is given below.
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
4.2K