Quantification of Metamorphopsia Using a Smartphone-Based Hyperacuity Test in Patients With Idiopathic Epiretinal
Daria Amon1, Christoph Leisser1, Andreas Schlatter1
1Vienna Institute for Research in Ocular Surgery, A Karl Landsteiner Institute, Hanusch Hospital, Vienna, Austria.
JMIR Perioperative Medicine
|April 17, 2025
Summary
A new smartphone-based hyperacuity test (SHT) effectively quantifies metamorphopsia in patients with idiopathic epiretinal membranes (iERMs). This tool shows significant improvement post-surgery, offering a practical solution for visual distortion assessment.
Area of Science:
- Ophthalmology
- Medical Technology
- Vision Science
Background:
- Idiopathic epiretinal membranes (iERMs) cause visual distortion (metamorphopsia), often underestimated clinically.
- Current assessment methods (visual acuity, OCT) inadequately reflect functional vision and metamorphopsia severity.
- A need exists for reliable tools to quantify metamorphopsia for improved patient outcomes.
Purpose of the Study:
- To assess a novel smartphone-based hyperacuity test (SHT) for quantifying metamorphopsia in iERM patients.
- To compare SHT performance with a conventional printed chart before and after surgery.
- To evaluate SHT's utility in clinical decision-making and telemedicine.
Main Methods:
- Prospective observational study of 27 iERM patients undergoing vitrectomy.
- Smartphone-based hyperacuity test (SHT) and M-chart tests performed pre- and post-surgery.
- Analysis of metamorphopsia scores, visual acuity, OCT biomarkers, and subjective perception.
Main Results:
- SHT scores significantly improved post-surgery (55.2 to 63.5, P<.001).
- M-chart scores showed insignificant improvement (MH P=.06, MV P=.36).
- Both SHT and M-chart demonstrated good reliability (ICC >0.75).
Conclusions:
- SHT is a valuable tool for quantifying visual distortion in iERM patients.
- SHT offers a practical, accessible solution for monitoring metamorphopsia, especially in telemedicine.
- The SHT can aid surgical decision-making and potentially reduce unnecessary patient visits.
Keywords:
AlleyeM-chartdistorted visionepiretinal membranehome monitoringhyperacuityhyperacuity testmacular degenerationmetamorphopsiamobile healthocularocular pathologyophthalmologyretinaretinalsmartphonesurgerysurgical interventiontelemedicinetomographyvisionvisual acuityvitrectomyvitrectomy with membrane peeling