Repeatability and agreement of three different ocular biometers including single and dual Scheimpflug cameras
Christoph Lwowski1, Klemens Paul Kaiser1, Kevin Ehrdorf1
1Department of Ophthalmology, Goethe-University, Frankfurt am Main, Germany.
Clinical Relevance:
With a wide range of commercially available biometry devices, direct comparisons for agreement, interchangeability and repeatability are needed to enable correct interpretation of measured values.
Background:
The purpose of this study was to evaluate the intra-device repeatability and inter-device agreement of three different ocular biometers.
Methods:
This is a prospective, comparative study conducted at the University Hospital, Goethe University, Frankfurt, Germany. Included were eyes with a cataract which showed no other pathology. One eye per patient was randomly selected and 3 consecutive measurements were taken with the IOLMaster 700, Galilei G6, and Pentacam AXL. Repeatability and agreement were evaluated for simulated keratometry (simK), astigmatism (simA) and its axis, total keratometry (TC), axial length (AL), anterior chamber depth (ACD), and pachymetry (CT).
Results:
simK, TC, AL, and CT show very high repeatability with an intra-class correlation (ICC) of > 0.9 and a COV < 0.5 each. Means of the measurements were significantly different between the devices for simK, AL, and TC (p < 0.05). ACD and simA also showed a high ICC, but a lower agreement between the devices. Fixed biases shown by Bland-Altman plots for simK seem to indicate that the values are not interchangeable.
Conclusion:
Even though very high repeatability and agreement for most of the measured variables was observed with all three devices, fixed differences indicate that measured values are not completely interchangeable between the devices.
More Related Videos
08:27Quantification of Oculomotor Responses and Accommodation Through Instrumentation and Analysis Toolboxes
Published on: March 3, 2023
06:19Comparison of Three Clinical Stereoscopic Methods for Measuring Binocular Visual Function During Amblyopic Treatment in Unilateral Amblyopia
Published on: September 27, 2024
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Confocal Fluorescence Microscopy
Three-Dimensional Microscopy in Microbiology