Identification and mitigation of a systematic analysis error in a multicenter dual-energy x-ray absorptiometry study
Gayle M Lorenzi1, Barbara H Braffett2, Ionut Bebu2
1University of California San Diego, La Jolla, CA, USA.
Clinical Trials (London, England)
|April 22, 2025
Summary
A systematic error in bone density scans led to incorrect osteoporosis diagnoses in a diabetes study. A quality assurance plan helped correct results and ensure participant safety, preventing unnecessary treatment.
Area of Science:
- Clinical Research
- Data Integrity
- Quality Assurance
Background:
- Multicenter and longitudinal studies require robust data integrity measures.
- Sharing results with participants necessitates accurate and reproducible data.
- A quality assurance plan is crucial for monitoring and mitigating errors.
Purpose of the Study:
- To report a systematic error in bone mineral density (BMD) hip scan analysis.
- To describe the mitigation and communication plan implemented to ensure participant safety.
- To highlight the importance of quality assurance in multicenter clinical studies.
Main Methods:
- Ancillary bone mineral density study across 27 sites using dual x-ray absorptiometry scanners.
- Identified a systematic error in hip scan readings from one scanner type.
- Implemented a mitigation plan involving scan reanalysis and participant communication.
Main Results:
- An error in hip scan analysis led to lower BMD scores, misclassifying osteoporosis.
- Fourteen scans initially indicating osteoporosis or low BMD were reanalyzed, with reclassifications occurring in 10 scans.
- A broader review identified additional scans requiring reanalysis, with 12 reclassifications.
Conclusions:
- Analysis errors can impact participant safety, potentially leading to unnecessary treatment.
- A coordinated communication and mitigation plan is essential for accurate results and participant care.
- This study provides a framework for managing data integrity issues in clinical research.
Related Concept Videos
Atomic Absorption Spectroscopy: Interference
478
Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
478
Systematic Error: Methodological and Sampling Errors
1.2K
In the case of systematic errors, the sources can be identified, and the errors can be subsequently minimized by addressing these sources. According to the source, systematic errors can be divided into sampling, instrumental, methodological, and personal errors.
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
1.2K