Reliability of hemodynamic parameters measured by bioimpedance cardiography at different intensities during
Raphael Schoch1, Jonathan Wagner1, Max Niemeyer2
1Department of Sport, Exercise and Health, Division Sports and Exercise Medicine, University of Basel, Basel, Switzerland.
Background:
Bioimpedance cardiography offers a non-invasive and time-efficient method to measure hemodynamic parameters. Previous studies only investigated its reliability under steady-state conditions and at maximum load but not at ventilatory thresholds (VTs). This is the first study that assesses the reliability of measured hemodynamic parameters at different exercise stages during cardiopulmonary exercise testing (CPET) using prespecified strict criteria to assess reliability.
Methods:
Data from 31 healthy, well-trained adults were analyzed. Each participant completed two CPETs, both following the same ramp protocol, with a 7-day interval between them. Hemodynamic parameters were measured with the PhysioFlow® (Manatec Biomedical, Poissy, France) at characteristic phases and thresholds [VT1, VT2, and peak oxygen uptake (V̇O2peak)]. To ensure comparability, the wattage (power) corresponding to the thresholds in Test 1 (PVT1, PVT2, and PV̇O2peak) was used for Test 2.
Results:
Heart rate, stroke volume, and cardiac output demonstrated good reliability on a group level (mean intraclass correlation >0.75) at both thresholds (0.91, 0.80, and 0.77 at PVT1; 0.92, 0.80, and 0.77 at PVT2) and at PV̇O2peak (0.93, 0.82, and 0.80). For stroke volume at PV̇O2peak, both individual differences (-39.0 to 36.9 mL for the women and -39.9 to 45.2 mL for the men) and mean detectable change (17.5 mL) were larger than the a priori defined acceptable ranges of agreement (-3.6 to 3.8 mL for the women and -4.5 to 3.3 mL for the men).
Conclusion:
The PhysioFlow® reliably measures heart rate, stroke volume, and cardiac output during CPET on a group level. However, as shown by the Bland-Altman plots, the reliability is too low to be used for individual comparisons.
Related Concept Videos
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...