Clinical Assessment of Automated Non-Contact Tonometer: Interchangeability with Goldmann Applanation Tonometry and
Michael Chaglasian1, Huiyuan Hou2, Mayra Tafreshi2
1Illinois College of Optometry, Chicago, IL 60616, USA.
The TRK-3 OMNIA non-contact tonometer (NCT) shows strong agreement and statistical equivalence with Goldmann applanation tonometry (GAT) for intraocular pressure (IOP) measurements. This automated tonometer offers reliable and repeatable IOP readings, supporting its use in clinical practice.
Area of Science:
- Ophthalmology
- Optometry
- Biomedical Engineering
Background:
- Accurate intraocular pressure (IOP) measurement is crucial for diagnosing and managing glaucoma.
- Goldmann applanation tonometry (GAT) is the gold standard, but non-contact tonometers (NCTs) offer convenience.
- Evaluating the clinical interchangeability and repeatability of new NCT devices is essential.
Purpose of the Study:
- To assess the clinical interchangeability of intraocular pressure (IOP) measurements between the TRK-3 OMNIA NCT and Goldmann applanation tonometry (GAT).
- To evaluate the repeatability of IOP measurements obtained with the TRK-3 OMNIA.
- To determine the correlation and agreement across different IOP ranges.
Main Methods:
- Prospective, multicenter study involving 120 subjects stratified into low, intermediate, and high IOP groups based on GAT.
- IOP measurements were taken using both TRK-3 OMNIA and GAT following a standardized protocol.
- Statistical analysis included Bland-Altman analysis, equivalence testing (TOST), and linear regression; repeatability assessed via ICC, repeatability limit, and CV.
Main Results:
- Mean difference between TRK-3 OMNIA and GAT was -0.2 mmHg, with strong correlation (r=0.92).
- Statistical equivalence confirmed across all IOP groups, despite slight overestimation in low and underestimation in high IOP ranges.
- TRK-3 OMNIA demonstrated good repeatability (ICC=0.94, CV=5.65%), with repeatability limits varying by IOP group.
Conclusions:
- The TRK-3 OMNIA tonometer shows strong agreement, statistical equivalence, and high correlation with GAT, supporting clinical interchangeability.
- The device exhibits high repeatability, indicating reliability for IOP measurements.
- The TRK-3 OMNIA is a suitable tool for routine clinical workflows, providing dependable IOP assessments.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...