Clinical Application of Monitoring Vital Signs in Dogs Through Ballistocardiography (BCG)
Bolortuya Chuluunbaatar1, YungAn Sun1, Kyerim Chang2
1College of Veterinary Medicine and Veterinary Medical Research Institute, Jeju National University, Jeju 63243, Republic of Korea.
Abstract:
This study evaluated the application of the BCG Sense1 wearable device for monitoring the heart rate (HR) and the respiratory rate (RR) in dogs, comparing its performance to the gold standard ECG under awake and anesthetized conditions. Data were collected from twelve dogs, with six awake beagles and six anesthetized client-owned dogs. Bland-Altman analysis and linear regression revealed strong correlations between BCG and ECG under both awake and anesthetized conditions (HR: r = 0.97, R2 = 0.94; RR: r = 0.78, R2 = 0.61, and p < 0.001). While slight irregularities were noted in respiratory rate measurements in both groups, potentially affecting the concordance between methods, BCG maintained a significant correlation with ECG under anesthesia (HR: r = 0.96, R2 = 0.92; RR: r = 0.85, R2 = 0.72, and p < 0.01). The wearable BCG-Sense 1 sensor enables continuous monitoring over 24 h, while ECG serves as the gold standard reference. These findings prove that BCG can be a good alternative to ECG for the monitoring of vital signs in clinical, perioperative, intraoperative, and postoperative settings. The strong correlation between the BCG and ECG signals in awake and anesthetized states highlights the prospects of BCG technology as a revolutionary method in veterinary medicine. As a non-invasive and real-time monitoring system, the BCG Sense1 device strengthens clinical diagnosis and reduces physiological variations induced by stress.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.