Evaluation of Two Veterinary Oscillometric Noninvasive Blood Pressure (NIBP) Measurement Devices (petMAP Graphic II
Hanna Walter1, Sabine B R Kästner1, Thomas Amon2
1Small Animal Clinic, University of Veterinary Medicine Hannover, 30559 Hannover, Germany.
Abstract:
Noninvasive blood pressure (NIBP) device performance in dogs may be influenced by extreme pressures and altered systemic vascular resistance (SVR). This study evaluated the agreement of two NIBP devices (HDO and petMAP) with invasive blood pressure (IBP) measurements, compliance with hypertension consensus statement criteria, and their trending ability (TA) across varying blood pressure and SVR ranges in awake and anesthetized dogs. Seven healthy Beagles were studied, with IBP recorded from the dorsal metatarsal artery and NIBP cuffs placed randomly on the front limb, hind limb, or base of the tail. Cardiac output was determined by thermodilution, and the systemic vascular resistance index (SVRI) was calculated by a standard formula. Bland-Altman, concordance rate, and polar plot analyses were used for statistical analysis. A total of 752 and 640 paired measurements were obtained for HDO and petMAP, respectively. Both devices showed good agreement with IBP for mean arterial pressure (MAP) at low blood pressure and the SVRI. At high blood pressure and the SVRI, agreement weakened, with substantial underestimation of systolic arterial pressure (SAP). Both devices demonstrated moderate to good TA for MAP and SAP. Overall, the best agreement was observed for MAP at a low SVRI, while agreement was moderate at hypertension (petMAP) and a high SVRI (petMAP, HDO).
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement