The Accuracy of a SmartBottle Device for Objective Assessment of Changes in Bottle Weight During Infant Feeding
Ellie Blach1, Molly M Ryder1, Robert Brewster2
1Department of Kinesiology and Public Health, Bailey College of Science and Mathematics, Center for Health Research, California Polytechnic State University, San Luis Obispo, California.
Background:
Current infant feeding assessment methods are often labor-intensive for caregivers and subject to bias.
Objective:
This study assessed the accuracy of the SmartBottle device, which attaches to an infant feeding bottle and automatically records the amount consumed during bottle-feeding.
Design:
This was a proof-of-concept study wherein changes in bottle weight during infant feeding interactions were simulated and compared with criterion weights. Research assistants conducted 4 trials to obtain SmartBottle-derived test weights and corresponding scale-derived criterion weights for bottles of various sizes, shapes, and formula volumes. Each trial focused on a different bottle: (1) Dr. Brown's 120 mL, (2) Dr. Brown's 250 mL, (3) Avent (Philips) 125 mL, and (4) Avent 260 mL. Twenty-five corresponding test and criterion weights were collected for each bottle, resulting in 100 measurements total.
Participants/Setting:
Three research assistants collected data within a controlled laboratory setting.
Main Outcome Measures:
SmartBottle-derived test weights and scale-derived criterion weights were the main outcome measures.
Statistical Analyses Performed:
Two 1-sided t tests were used to compare the equivalence of SmartBottle-derived test weights to scale-derived criterion weights at ±5%, ±7.5%, and ±10% of scale-derived criterion weights. Bland-Altman plots were examined to assess whether there were systematic errors in SmartBottle-derived test weights.
Results:
The mean difference between SmartBottle-derived test weights and scale-derived criterion weights was -0.01 (90% CI -0.02 to 0.00) and was equivalent within ±5% bounds (P < .0001). Bland-Altman plots illustrated that there was no evidence of systematic bias related to bottle weight, size, or shape.
Conclusions:
SmartBottle-derived test weights were equivalent to scale-derived criterion weights within ±5% equivalence bounds with no evidence of bias. Further testing within real-world infant feeding contexts is warranted.
More Related Videos
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Uncertainty in Measurement: Reading Instruments
Uncertainty in Measurement: Accuracy and Precision
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...