Hemodynamic Monitoring During Hemodialysis Using Bioimpedance: A Comparison of Changes in Resistance Between
Melanie K Schoutteten1,2,3, Patrick van der Heijden3, Astrid D H Brys4
1Hasselt University, Faculty of Biomedical and Life Sciences, Limburg Clinical Research Center/Mobile Health Unit, Diepenbeek, Belgium.
Introduction:
Prevention of hemodynamic complications during hemodialysis remains challenging. Although whole body bioimpedance is well established in fluid status assessment, its use for dynamic or continuous recordings is limited. A segmental approach may serve this purpose better. This study investigates which body segment is best targeted to measure bioimpedance for hemodynamic monitoring.
Methods:
In this observational study, serial bioimpedance measurements were conducted on the whole body, lower leg, upper arm, and thorax of 15 patients during two hemodialysis sessions. The resistance component of bioimpedance was used to investigate the relationship with changes in volume and systolic blood pressure (SBP).
Findings:
Predialysis to postdialysis changes in relative resistance between the two sessions revealed the lowest intraclass correlation coefficient for upper arm (0.023) and the highest for thoracic resistance (0.728). Correlation between ultrafiltration volume and relative resistance was comparable between upper arm and thoracic segment (0.538 [0.447-0.618] and 0.537 [0.446-0.617], both p < 0.001, respectively) and the highest for whole-body and lower leg (0.697 [0.63-0.754] and 0.670 [0.598-0.731], both p < 0.001, respectively). In contrast, the correlation between changes in SBP and relative resistance was the highest in the thoracic segment (-0.33 [-0.432 to -0.219], p < 0.001) and the lowest for whole body measurements (-0.154 [-0.269 to -0.036], p = 0.01). In addition, multiple regression analysis indicated thoracic resistance as the best predictor for changes in SBP (β = -0.261 [-0.353 to -0.126], p < 0.001).
Discussion:
These findings suggest that the thorax is the most suitable region for segmental bioimpedance measurements to assess hemodynamic parameters. Thoracic bioimpedance may innovate the hemodynamic monitoring of hemodialysis patients.
Related Concept Videos
Dialysis
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...