Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
First-in-human study of a leadless pacemaker system for left bundle branch area pacing.
Vivek Y Reddy1, Devi G Nair2, Shephal K Doshi3
1Helmsley Electrophysiology Center, Mount Sinai Fuster Heart Hospital, New York, New York; Na Homolce Hospital & Charles University, Prague, Czech Republic.
This study shows a new leadless pacemaker (LPCSP) can be safely implanted for left bundle branch area pacing (LBBAP). The novel device demonstrated acceptable electrical performance and safety in a first-in-human trial.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Leadless pacemakers (LPs) currently cannot perform left bundle branch area pacing (LBBAP).
- LBBAP is a promising pacing strategy for certain cardiac conditions.
Purpose of the Study:
- To evaluate the safety and performance of a novel helix-based LP with an extended electrode for LBBAP (LPCSP).
- This was a first-in-human acute feasibility study.
Main Methods:
- The LPCSP was temporarily deployed in patients undergoing LP implantation via the right internal jugular vein.
- Implantation into the interventricular septum (IVS) was guided by fluoroscopy and intracardiac echocardiography.
- Primary endpoints included freedom from serious adverse events and intraprocedural electrical parameters.
Main Results:
- The LPCSP was successfully implanted in 10 out of 14 patients.
- Acceptable pacing thresholds, R-wave amplitudes, and impedance were recorded.
- Left bundle branch area pacing (LBBAP) capture was achieved in 5 patients, with others showing deep septal pacing.
Conclusions:
- The novel LPCSP demonstrated acute successful implantation.
- The device showed clinically acceptable intraprocedural electrical performance and safety.
- This study supports the feasibility of using this device for LBBAP.