Related Experiment Video
Updated: May 9, 2025

12:08
Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
18.5K
Real-World Accuracy of a Continuous Glucose Monitoring System after Radiologic Exposure
Siobhan Tellez1, Lindsey Hornung2, Emily Smith1
1Division of Endocrinology Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA.
Pediatric Diabetes
|April 30, 2025
Summary
Continuous glucose monitor (CGM) accuracy is unaffected by scatter radiation during diagnostic imaging. Leaving CGMs in place during X-ray or CT scans reduces patient burden and costs.
Area of Science:
- Endocrinology
- Medical Imaging
- Biomedical Engineering
Background:
- Continuous glucose monitor (CGM) use is increasing, but manufacturer guidelines recommend removal during diagnostic imaging (X-ray, CT).
- This removal imposes financial, clinical, and psychosocial burdens, particularly for pediatric patients undergoing total pancreatectomy with islet autotransplantation (TPIAT).
- Interruptions in CGM use can hinder optimal diabetes management.
Purpose of the Study:
- To evaluate the impact of scatter radiation dose exposure on CGM accuracy.
- To assess if leaving CGMs in place during diagnostic imaging, outside the field of view, affects accuracy.
- To determine if this practice mitigates interruptions and burdens for TPIAT patients.
Main Methods:
- 14 pediatric TPIAT patients using insulin pumps and CGMs were monitored for 3 months post-surgery.
- CGMs were unshielded during routine X-ray or CT procedures.
- Absolute differences between CGM and glucometer values were analyzed using mixed models; Clarke error grid analyses (EGA) were performed.
- Scatter dose exposure was calculated retrospectively using anthropomorphic phantoms.
Main Results:
- Patients received a median of five imaging procedures with a median cumulative scatter dose of 559 µGy.
- No significant association was found between scatter dose exposure and the absolute difference between CGM and glucometer values (p=0.17).
- Over 98% of glucose values fell within clinically acceptable zones (A and B) on EGA, irrespective of scatter dose or time from TPIAT.
Conclusions:
- Scatter dose exposure from diagnostic imaging does not clinically impact CGM accuracy.
- Leaving CGM components in place when outside the field of view during imaging is safe and effective.
- This practice reduces interruptions, burdens, and costs for patients, improving diabetes management.
Related Concept Videos
Biological Effects of Radiation
15.0K
All radioactive nuclides emit high-energy particles or electromagnetic waves. When this radiation encounters living cells, it can cause heating, break chemical bonds, or ionize molecules. The most serious biological damage results when these radioactive emissions fragment or ionize molecules. For example, α and β particles emitted from nuclear decay reactions possess much higher energies than ordinary chemical bond energies. When these particles strike and penetrate matter, they...
15.0K
Hypoglycemia and Glucagon
134
Without prolonged fasting, healthy individuals maintain blood glucose levels above 3.5 mM due to a well-adapted neuroendocrine counterregulatory system that effectively prevents acute hypoglycemia, a potentially life-threatening condition. The primary clinical scenarios for hypoglycemia encompass diabetes treatment, inappropriate production of endogenous insulin or insulin-like substances by tumors, and the use of glucose-lowering agents in non-diabetic individuals. Notably, hypoglycemia in the...
134
Imaging Studies II: Positron Emission Tomography and Scintigraphy
39
Positron Emission Tomography (PET) is a medical imaging technique that provides crucial insights into the body's physiological functions at a molecular level. It is an indispensable resource for diagnosing, staging, and monitoring various illnesses, notably cancer, neurological disorders, and cardiovascular conditions.
Fundamental Principles of PET
Fundamental Principles of PET
39

