Clinical evaluation of two methods of subdural pressure monitoring
Abstract:
Recordings from two different types of subdural pressure monitor with simultaneous intraventricular pressure (IVP) tracings are compared in 20 head-injured patients. In the first 10 patients a fluid-filled catheter was placed subdurally and connected to an external transducer, and in the second 10 the Gaeltec model ICT/b solid state miniature transducer was used. The latter system has the advantage that both zero and calibration checks can be carried out after insertion. Only 44% of the fluid-filled catheter readings corresponded with IVP in series of 10-mm Hg ranges, while 53% of readings were lower; this tendency was more marked at higher pressures. With the Gaeltec transducer, 72% of subdural pressure readings corresponded with IVP, while only 9% were lower and 19% were higher than IVP. The differences may have been due to technical causes or to true pressure differentials. The subdural catheter appears too unreliable for routine clinical use, but the Gaeltec transducer may be a satisfactory alternative to ventricular pressure monitoring.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.