Eléonore Vermeulen1, Pierre-Yves Baudin1, Marc Lapert2
1NMR Laboratory, Neuromuscular Investigation Center, Institute of Myology, Paris, France.
Magnetic Resonance in Medicine
|May 6, 2025
Summary
A new 3D MRI sequence accurately estimates water T2 in skeletal muscles, even with motion. This method improves characterization of muscle tissue and is suitable for clinical research.
Area of Science:
- Magnetic Resonance Imaging
- Biomedical Engineering
- Musculoskeletal Imaging
Background:
- Water T2 relaxation time is a crucial biomarker for skeletal muscle tissue characterization.
- Accurate T2 estimation is challenging in skeletal muscles due to motion and B1 field inhomogeneities.
- Existing methods often lack the speed and robustness required for in vivo applications.
Purpose of the Study:
- To develop and validate a motion-robust 3D magnetic resonance imaging (MRI) sequence for accurate water T2 estimation in skeletal muscles.
- To assess the sequence's performance in phantoms and in vivo, including its ability to account for confounding factors like B1 variations and fat content.
Main Methods:
- A partially spoiled gradient echo (pSPGR) sequence was employed, acquiring 10 image volumes with varied RF phase-cycling and flip angles.
- A bi-component water/fat model was used to fit the complex signal evolution, extracting T2 values while correcting for B1 and fat fraction.
- The sequence was implemented in both Cartesian and radial trajectories, with validation through numerical simulations, phantom studies, and in vivo experiments on thigh and tongue muscles.
Main Results:
- Phantom studies showed high correlation (R² > 0.8) between the proposed method and reference spectroscopy/multi-spin echo techniques.
- In vivo, T2 values correlated well with reference methods in healthy (R² = 0.69) and pathological muscles (R² = 0.87), demonstrating robustness to B1 inhomogeneities (R² = 0.06).
- The radial implementation significantly reduced the standard deviation of T2 values in the tongue muscle by 28% compared to the Cartesian approach.
Conclusions:
- The proposed 3D sequence enables efficient and accurate water T2 estimation in skeletal muscles, including small, moving structures like the tongue.
- This technique expands the potential for characterizing heterogeneous muscle impairments.
- The method's compatibility with clinical research durations makes it a valuable tool for muscle tissue assessment.