Artifact-Tolerant Electrophysiological Sensor Interface With 3.6V/1.8V DM/CM Input Range and 52.3mVpp/${\mu}$s
Abstract:
In the practical implementations of wearable sensors, motion artifacts with large amplitudes often cause signal chain saturation, significantly degrading biopotential signal integrity. Similarly, rapid stimulation artifacts are inevitable during closed-loop brain stimulation therapy, posing additional challenges for real-time signal acquisition. To address motion and stimulation artifacts with amplitudes reaching hundreds of mV while minimizing information loss, a sensor interface with high input range and fast artifacts recovery capability is essential. This paper presents a continuous-time track-and-zoom (CT-TAZ) technique designed to handle large artifacts events without saturation. The proposed system achieves a 3.6V/1.8V differential-mode/common-mode full-scale input range. Fabricated in a 180nm CMOS process, the prototype chip occupies an area of 0.694mm2 and consumes 12/32.6/51.6$\mu$W for recordings without/with single-end/with differential rail-to-rail artifacts. The system demonstrates an average artifacts recovery time of 65.3 $\mu$s under 3.6V stimulation artifacts, achieving an average artifacts recovery speed of 52.3mVpp/μs, which is 2.25× larger input range and 3× faster recovery compared to the state-of-the-art.