PVI With CF-Sensing Large-Tip Focal PFA Catheter With 3D Mapping for Paroxysmal AF: Omny-IRE 3-Month Results
Mattias Duytschaever1, Massimo Grimaldi2, Tom De Potter3
1Department of Cardiology, AZ Sint-Jan Hospital, Bruges, Belgium.
JACC. Clinical Electrophysiology
|May 9, 2025
Summary
The OMNYPULSE pulsed field ablation catheter achieved 100% acute success in treating paroxysmal atrial fibrillation (PAF). This novel catheter demonstrated a promising safety profile and good pulmonary vein isolation durability in the Omny-IRE study.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- The Omny-IRE study evaluated the OMNYPULSE catheter, a novel large-tip, multielectrode, contact force-sensing pulsed field ablation device integrated with electroanatomic mapping.
- This technology aims to improve the treatment of paroxysmal atrial fibrillation (PAF).
Purpose of the Study:
- To assess the 3-month safety and effectiveness of the OMNYPULSE platform for treating symptomatic paroxysmal atrial fibrillation.
- To evaluate the durability of pulmonary vein isolation (PVI) achieved with the novel catheter system.
Main Methods:
- Pulmonary vein isolation (PVI) was performed using the OMNYPULSE Platform.
- Primary effectiveness was defined as adenosine/isoproterenol-proof entrance block.
- Primary safety was assessed by the occurrence of primary adverse events, with prespecified subsets undergoing brain imaging, esophageal endoscopy, and cardiac imaging.
Main Results:
- 100% acute effectiveness (136/136 patients) was achieved.
- The primary adverse event rate was 3.0% (4/135 patients), including vascular access complications and pericarditis.
- Pulmonary vein isolation (PVI) durability was 62.1% at 3 months, improving to 71.4% with an optimized workflow.
Conclusions:
- The force-sensing, large-focal pulsed field ablation catheter with integrated 3D electroanatomic mapping demonstrated 100% acute success for paroxysmal atrial fibrillation treatment.
- The platform exhibited a promising safety profile.
- Prespecified remapping indicated good PVI durability, suggesting potential for effective long-term treatment.