The Hidden Threat in Immunoassays: A Case of Persistent False-Positive CK-MB
Siwei Xue1, Weiqiang Huang2, Degui Ou3
1Department of Laboratory, Huiyang Sanhe Hospital, Huizhou, China.
Objective:
This report aims to explore the significant impact of heterophilic antibodies on CK-MB immunoassays, leading to persistent false-positive results.
Case Report:
A 59-year-old male with a history of coronary artery disease exhibited persistently elevated CK-MB levels for over two years, despite normal troponin I levels. Further investigations, including serial dilution, blocking experiments, and testing across different platforms, confirmed the presence of heterophilic antibody interference, which was responsible for the falsely elevated CK-MB levels.
Conclusion:
This case underscores the critical importance of recognizing heterophilic antibody interference in immunoassays. The use of appropriate diagnostic approaches, such as blocking antibodies, cross-platform testing, and the integration of additional diagnostic markers, is essential to prevent misdiagnosis and ensure accurate clinical decision-making.