Related Experiment Video
Updated: May 8, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Radiation exposure in XLIF surgery utilizing ultra-low radiation imaging with image enhancement software: a
Benno Bullert1, Jula Gierse1, Eric Mandelka1
1BG Klinik Ludwigshafen, Department of Orthopaedics and Trauma Surgery, Heidelberg University, Research Group Medical Imaging and Navigation in Trauma and Orthopaedic Surgery (MINTOS), Ludwig-Guttmann-Str. 13, Ludwigshafen 67071, Germany; Heidelberg University, Grabengasse 1, Heidelberg D-69117, Germany.
Background Context:
Minimally invasive spine surgery has grown significantly over the last decade. Lateral approaches for interbody fusion, such as XLIF/LLIF result in higher radiation exposure for the operating room (OR) staff and the patient compared to posterior techniques. New technologies such as ultra-low radiation imaging (ULRI) combined with image enhancement (IE) software may help reduce this exposure while maintaining procedural efficacy.
Purpose:
To evaluate whether using ULRI with IE software (mFluoro) reduces radiation exposure for OR staff and patients during lateral interbody fusion via the XLIF approach without increasing procedure time.
Study Design:
Prospective, randomized controlled trial.
Patient Sample:
The sample consists of 60 patients.
Outcome Measures:
Primary outcome: Radiation exposure of the OR staff, measured in microsieverts (μSv), documented by personal dosimetry centrally in front of the sternum above the lead apron.
Secondary Outcomes:
Radiation exposure of the patient, measured in cGy*cm2, documented by the fluoroscopy unit's dose report; Procedure time, measured in minutes.
Methods:
This single-center prospective randomized controlled single-blind study included 60 patients who all underwent single-level lateral interbody fusion via XLIF approach between March, 2023 and December, 2024. Patients were randomized into 2 groups: intervention group (mFluoro) or control group (cFluoro). Radiation exposure of the OR staff was measured using dosimeters, and imaging parameters were extracted from the fluoroscopy unit's dose report.
Results:
No significant difference was found between the 2 groups regarding age, gender, BMI, implant type, and diagnosis (all p>.05). The mFluoro group showed a significant reduction in radiation exposure for the OR staff: 72.1% for the surgeon (p<.001), 76.1% for the assistant (p<.001), and 67.5% for the scrub nurse (p<.001). Patient radiation exposure was also significantly reduced in the intervention group, with dose area product (DAP) values lowered by 66.1% (p<.001). Fewer images were acquired in the mFluoro group (31.3% reduction, p<.001), and procedure time was reduced by 17.7% (71.3±21.3 minutes vs 86.6±31.5 minutes; p=.034).
Conclusion:
Using ULRI with IE software significantly reduces radiation exposure for the OR staff in single-level lateral interbody fusion via the XLIF approach. Furthermore, the radiation exposure for the patient was also significantly reduced.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...