Related Experiment Video
Updated: May 15, 2025

Effects of Surgical Masks on Cardiopulmonary Function in Healthy Subjects
Published on: February 12, 2021
Particle filtration efficiency of respirators during the COVID-19 pandemic
Jackson Williams1, Lucan Baillie1, Wayne Clancy1
1Health Products Regulation Group, Therapeutic Goods Administration, 27 Scherger Drive Fairbairn, Australian Capital Territory 2609, Australia.
Background:
Respirators are an essential medical device vital to prevent the transmission of airborne illness between persons. This study presents an investigation conducted by the Therapeutic Goods Administration (TGA) into the particle filtration efficiency (PFE) performance of respirators supplied in Australia that were included as medical devices in the Australian Register of Therapeutic Goods (ARTG) throughout the COVID-19 pandemic or supplied to the TGA from other jurisdictions for compliance testing.
Methods:
A total of 400 real-world batches of respirators were tested against a rapid-screen PFE method developed by the TGA. The TGA analysed a respirator's ability to meet its claimed filtration efficiency with a statistical power analysis differentiated by the claimed PFE standard, country of manufacture, time of inclusion into the ARTG and respirators with surgical against non-surgical claims.
Findings:
Of the tested batches of respirators (n = 400), 70 % were compliant with the pre-set compliance criteria. The results indicate statistically significant PFE differences (p < 0.05) between respirators claiming the standards 42 CFR Part 84, AS 1716:2012, GB 19083: 2010 and EN 149:2001+ A1:2009 when tested against respirators claiming GB 2626:2006. There were no observed significant differences between the PFE of surgical and non-surgical respirators (p > 0.05).
Interpretation:
The findings of this study highlight the importance of ensuring respirators meet their claimed PFE level based on their claimed standard prior to manufacturer release. Further studies should be focussed on investigating the effect of respiratory protection devices and their claimed efficiency in clinical settings such as hospitals against aerosolised particulates using a range of respiratory protection devices.
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Personal Protective Equipment
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: