Comparison of Visual Performance Between Two Diffractive Trifocal Intraocular Lenses
Gloria Segura-Duch1, David Oliver-Gutierrez1, Mar Arans1
1Verte-ICO Oftalmología, 08006 Barcelona, Spain.
Abstract:
Background/Objectives: This study aimed to compare the visual outcomes of two diffractive trifocal intraocular lenses (IOLs): the Bi-Flex Liberty 677MY (Medicontur; Zsámbék, Hungary) and the FineVision POD F (BVI Medical; Waltham, MA, USA). Methods: A prospective study with a 3-month follow-up was conducted. A total of 62 patients were divided into two groups according to the type of lens implanted: 31 patients with the Liberty lens (61.1 ± 6.4 years) and 31 patients with the Finevision lens (61.9 ± 6.8 years). Three measurement sessions were conducted (baseline, 1 month, and 3 months). These sessions included measurements of the subjective refraction, visual acuity, and defocus curve. Both eyes of each patient were operated on and included in the statistical analysis. Results: Three months after surgery, monocular visual acuity with the Liberty lens was statistically greater than with the Finevision lens at defocus values of -2.00 D (50 cm) and -2.50 D (40 cm) (p < 0.01). In this regard, the near visual acuity results (40-50 cm) with the Liberty lens showed greater variability compared to those of the Finevision lens. Binocularly, however, the Finevision lens demonstrated a statistically significant improvement in visual acuity than the Liberty lens at a defocus of -1.50 D (67 cm) (p = 0.01). Both IOLs provided visual acuities better than 0.20 logMAR for a defocus range from distance (0.00 D) to near (-3.50 D). Conclusions: Future studies are needed to investigate which patient ocular parameters could predict improved near vision with the Liberty lens or intermediate vision with the Finevision lens.
Related Concept Videos
Focusing of Light in the Eye
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...