1Department of Restorative Dental Sciences, College of Dentistry, Imam Abdulrahman Bin Faisal University, P.O.Box 1982, Dammam, 31441, Saudi Arabia. abalhaddad@iau.edu.sa.
Objective:
The light-curing unit (LCU) has become a vital piece of dental equipment that must be correctly maintained. This study investigated the impact of contamination and physical damage to the light tip on the power and radiant emittance values from old and new LCUs.
Materials And Methods:
Two investigators assessed 200 LCUs in dental clinics. The extent of contamination and physical damage to the light-curing unit (LCU) tips was recorded using a scale ranging from 0 to 8, where 0 indicates the absence of damage or contamination, and 8 represents severe damage or contamination. Then, the radiant emittance and power values of the LCU tip were measured using a digital radiometer (Bluephase meter II; Ivoclar, Schaan, Liechtenstein). LCUs that were more than five years old were classified as old. Spearman correlation coefficient was used to determine the relationship between the condition of the LCU and radiant emittance/power (p = 0.05).
Results:
There were no significant differences in the percent reduction of the power and radiant emittance from the values reported by the manufacturers, as well as the presence of contamination or physical damage scores between old and relatively new light-curing tips (p > 0.05). The mean ± standard deviation percentage reductions in power and radiant emittance from the manufacturer's stated values were 19.2 ± 17.63% and 3.9 ± 16.49%, respectively. Contamination and physical damage had significant positive correlations with the reduction in the power (r = 0.22070, p = 0.0017 and r = 0.27422, p < 0.0001, respectively) and the reduction in the radiant emittance (r = 0.28626, p < 0.0001 and r = 0.36650, p < 0.0001). Increased contamination and physical damage scores corresponded to greater percent reductions in the power and radiant emittance (p < 0.05).
Conclusions:
Contamination and physical damage to the LCU can negatively impact the light output from LCUs.
Clinical Relevance:
To ensure optimal performance, dentists should regularly monitor the output of their LCUs and examine the devices for any signs of physical damage or contamination.